What is the difference of the vanceometer to test a variety of different angles





Vanceometer is designed to measure a variety of coating a surface gloss, such as ceramics, paints, inks, plastics, marble, aluminum, metal and other material surface gloss. Is a product surface gloss control instrument. In general the test, you can choose 40 °, 60 ° and 85 °angle. Do you know these three angle test has any difference? We have to choose what kind of angle can be most suitable for our use?
The Gloss (Universal angle): 60 °
60 ° is applied to all surfaces measurement angle. It is also used as a low-gloss 85 °, high-gloss 40 °is reference angle of the measure surface.

Low Gloss: 85 °
In order to improve the resolution of lowgloss surface. Available 85 ° gloss measurement, when measured with a 60 ° surface gloss values below 10GU, recommended to use 85 ° angle measurement.

When using the 85 ° gloss measurement of different texture or slightly uneven surface, the need for multi-point measurements averaged.

High Gloss: 40 °
40 ° measuring angle can improve the resolution of a highgloss surface. When measured with a 60 ° higher than the object surface brightness value 70GU, it recommended to use 40 ° angle measurement.

40 ° angle (light path) of the surface haze phenomenon is very sensitive. Because the haze is peculiar phenomenon of  high-gloss surface.

2015年11月22日星期日

The features of Colorimeter


1. The automatic comparison of the model with the color difference between the seized goods, the output CIE_Lab three sets of data and colorimetric after △ E, △ L, △ a, △ b four sets of color data, providing color reference solution.

2. With a single sample and two measurement modes, to meet the needs of different places measurements, simple operation, measurement precision.

3. The instrument is portable. Battery and external power supply two power supply, convenient and practical.

4. Equipped with USB expansion interface can be connected to the computer display. Colorimeter, widely used in the field with color management of plastic , printing, ink painting, textile and apparel industries.


Torque tester


HC―NJ Static torque tester applicable to many kinds torsional test and test for adjusting the to

rsional destroy parts. Mainly use for all kinds of electric,pneumatic torque approved,torque scre

wdriver , screwdriver,torque wrench and a variety of tighten force instruments and products,Wid

ely use in electrical manufacturing , machinery manufacturing,automotive light industry and prof

essional research and testing industry.
★★★★★★★★★★★★★★★★★★★★★★★★★★★
Feature
1.LED double figures with backlit display.
2.USD data interface,support hot-plug.
3.Set the max and min deviation,and free to select the alarm mode within limits and external limit.
4.Three kind of units N.m, kgf.cm , lbf.iN can be free conversion,and also can automatic identify twist direction.
5.With peak keeping function,the peak automatically discharge and time discharge can be set freely.
6.It can permanently store 10groups test reports and automatically calculate out the peak,valley,average and etc in the test report
7.The power―saving function shutdown automatically whiles no operation and automatically backlight.
8.Use the specified software on the PC can store,deliver,review and print the test curve data.


2015年11月15日星期日

Plastic testing should have the basic instrumentation

Plastics (including plastic products, plastic raw materials, particles) which are necessary testing equipment has it? Let us enumerate some of the commonly used plastic testing instruments, (for reference)
♠♣♦♣♠♦♣♠♦♠♣♦♣♠♦♣♠♦♠♣♦♣♠♦♣♠♦♠♣♦♣♠♦♣♠♦♠♣♦♣♠♦♣♠♦♠♣♦♣♠♦♣♠♦♠♣♦♣♠♦♣♠♦♠♣♦♣♠♦
1.Melt flow rate tester: for the determination of various plastics, resin melt flow rate MFR value when the viscous flow state, both suitable for high melting temperature polycarbonate, poly aromatic sulfone, fluorine plastic, nylon and other engineering plastics, are also suitable for polyethylene (PE), polystyrene (PS), polypropylene (PP), ABS resin, polyoxymethylene (POM), polycarbonate (PC) resin, a low melting temperature plastic test . Meet the standards: ISO 1133-1997, ASTM D1238-04C, GB / T3682-2000
◆♦◆♦◆♦◆♦◆♦◆♦◆♦◆♦◆♦◆♦◆♦◆♦◆♦◆♦◆♦◆♦◆♦◆♦◆♦◆♦◆♦◆♦◆♦◆♦◆♦◆♦◆♦
2.Charpy (Izod) impact tester: for the determination of plastic toughness, and impact resistance of the instrument 
3. Densitometer: for testing the density of plastic.
4. HTD&vicat softening point temperature tester: Softening point test for plastics.
5.Electronic universal testing machine: for physical performance testing of plastics, such as: tensile strength, flexural strength, elongation, etc.
6.Low temperature brittleness tester: impact test for the detection of plastics at low temperatures.
7.Dumbbell sample milling machine: a standard sample used for the production of plastics (injection molding machine can also be completed)
●★●★●★●●★●★●★●●★●★●★●●★●★●★●●★●★●★●●★●★●★●
These are plastic testing should have the basic instrumentation, more product information and detailed technical data, please visit the company website  www.gzxlyq.com or contact us Buyers

2015年11月6日星期五

ETB-068 Vancemeter

Product specification

ETB -068 series mirror, its gloss measurement instrument technology parameters are completely accord with national standard GB/T9754-88 GB/T9966.5 , international standard of ISO2813, each index reaches the national JJG696-2002 (mirror glossiness instrument calibration procedures) level of job of requirements.
♦ . ♦ .  ♦ . ♦ . ♦ . ♦ . ♦ . ♦ . ♦ . ♦ . ♦ . ♦ . ♦ . ♦ . ♦ . ♦ . ♦ . 

Scope of application

Paint, ink, coating, wood surface gloss measurement; Building decoration materials, marble, granite, polishing brick, ceramic brick surface gloss measurement; Plastic, paper and other surface gloss measurement, Other nonmetal material surface gloss measurement.
♦ . ♦ .  ♦ . ♦ . ♦ . ♦ . ♦ . ♦ . ♦ . ♦ . ♦ . ♦ . ♦ . ♦ . ♦ . ♦ . ♦ . 






Instrument characteristics

Fine, lightly appearance, easy to carry
The intelligent design, one-touch operation,easy to use
Various angles, any choice
Long-life light, no replacement
Measurement accuracy, repeatability
LCD digital display
Operating process has hum
Has the power supply voltage instruction function
Random attach a charger,two rechargeable batteries
Has the function of automatic shutdown
Technical parameters:
Testing range:0-200Gs
Stability:±0.4Gs/Min
Values error:±0.2Gs
Power:9v
Angle projection:20°  60°  85°
Environmental temperature:0-40degree
Relative humidity: Don't exceed 85%
Standard plate size:95×40MM


2015年11月1日星期日

High and low temperature alternating temperature humidity test chamber common troubleshooting method of the fault.

1. In doing damp heat test, the emergence of the actual humidity will reach 100% or actual humidity and the target humidity vary greatly, a lot of low value, the former phenomenon: It may be gauze wet bulb sensor caused by dry, wet it would have to check Sink the ball sensor whether water, the water level in the tank is automatically controlled by a water level controller, the pipeline groove check whether the water level controller water system if the normal water level controller is working properly. Another possibility is the use of wet autumn gauze long time, or the quality of water purity reasons, gauze will harden, so can not absorb moisture and dry gauze, if you replace or clean gauze to eliminate the above phenomenon. The latter phenomenon is the humidification system is not working, view the humidification system, water supply system, if there is a certain amount of water inside the water supply systems, boiler water level control humidification level control is normal, if normal water level inside the humidifier. Magnetic pump as above normal, it would have to check the electrical control system, but also to please the professional service personnel.

2. In the high_temperature tests, such as temperature changes reach the test temperature, you can check the electrical system, one by one to troubleshoot. As the temperature rises very slowly, it is necessary to view the air circulation system, look at the wind circulation damper is open normally, otherwise, it is necessary to check the motor running air circulation is normal. Sewage pump as temperature overshoot badly then you need to set the parameters tuning PID. If the temperature rises directly, overtemperature protection, then the controller fails, replace the control instrument.


3. The low temperature reach indicator test, then you have to observe the change in temperature, the temperature drop is very slow, or the temperature to a certain value temperature tends to increase, the former is necessary to check, whether the money will do the low temperature test Studios drying, the studio will remain dry before the test sample into the studio to do the test, the test sample too much work if placed indoors, chemical pumps so that indoor air can not fully cycle, after the exclusion of the above reasons , we must consider whether the refrigeration system broke down, so we must ask the manufacturer for repair professionals. The latter phenomenon is due to poor use of environmental equipment, the device is placed in the ambient temperature, place the position (distance from the wall after the box) to meet the requirements (in the manual operation of the test device has specified)

4. Device sudden failure during the test operation, the control corresponding fault appears on the instrument display and a voice prompt alarm. The operator can control the operation of equipment used in the troubleshooting chapter quickly check which type fault, you can ask professionals quickly troubleshoot to ensure proper conduct testsOther environmental testing equipment in use, there will be other phenomena, it should be specific phenomenon, detailed analysis and troubleshooting.
Environmental testing equipment but also regular maintenance condensers, cooling system cleaned regularly, shall be specified for lubrication of moving parts, electrical control systems periodic maintenance inspection, the work is essential.


Tensile testing machine require conditions and predictability in the future

With the development of machinery and equipment, green energy is particularly important, tensile testing machine also will embark on the road of green, energy saving green development on the roadconstantly, the future direction of development:
1, higher operating speed;
2, getting longer and longer service life;
3, effectively save energy and costs;
4, effectively reduce the operating temperature, reduce energy consumption;
5, based on the use of environmentally acceptable lubricants heavier loads;
Tensile testing machine need to have condition:
1) Test rubber requires a large stroke
When the rubber tensile deformation is large, especially latex products, elongation up may be to 1000% or more . So before rubber specimen fracture, tensile testing machine must ensure rubber holder has sufficient stroke.
2) high-precision and high-frequency data collection
Stretching the rubber does not require great force, pulling a large measurement range is not required, so the high accuracy required force value. General requirements for testing machines can strike more than two decimal places precision force values. Xiangli instruments precision value of rubber tensile testing machine reaches ± 0.5%. Furthermore, since the detection of the tensile properties of rubber stretching process required several tension value, tensile test and can not be repeated, so immediate and accurate record of each test section tensile strength values ​​for the test plays a very important role in the success.
3) accurate gauge measuring and recording devices.
Measuring gage length is calculated rubber elongation of important data,so tensile testing machine must accurately measure the strain of the sample in the rubber tensile test, and instantly recorded.
4) can accurately describe the stress - strain curve of the device.
Tensile specimens in tension between values and gauge closely linked, for example: modulus tensile specimen sample to be measured to a given elongation force value, depending on the stress you need to measure sample stretched to a given stress gauge