Measurement as follow:
The position of the measuring point: the measurement point is located in front of the device axis 1m distance away (device front vertical), the height from the ground to a height of 1/2 of the device, but the maximum height from the ground is not more than 1.5m, the minimum height of not less than 1m.
Measurement:
1, before the open the temperature &humidity test chamber, measuring measurement point A‐weighted sound pressure level of the background noise.
2, the device under test load and maximum radiated noise working conditions normal and stable operation, the use of the A‐weighted network measurements level and A‐weighted sound pressure level , the microphone should be positive for the test equipment, the use of sound‐level meter "slow "time weighting characteristics were measured, sound pressure level is the average count within the observation period (for maximum or minimum occasional not be considered). To avoid affecting the operation's body reflection measurements, the operating distance microphone should be at least greater than 0.5m.
3. Record the measured values, according to Table 1 as amended, temperature & humidity test chamber is running noise A-weighted sound pressure level.